STRIV next to two lab references: a Bertec force-plate treadmill and a Footscan pressure plate.
Gold-standard validated.
Force-plate treadmills are the gold standard for measuring running forces in biomechanics labs. STRIV tracked the same changes as Bertec across ground reaction force (GRF), contact time, vertical average loading rate (VALR), loading time, and braking force.
Validated over 40,000 steps · 24 runnersHow we tested
Runners changed speed on a Bertec force-plate treadmill. Each point averages 15 steps. Average change error compares changes from each recording’s usual level. BW means body weight; BW/s means body weights per second.
Choose a measurement to see its average change error.
Average change error · lower is closer to the lab.
Peak vertical GRF
Lab-plate pressure, inside the shoe.
We ran STRIV head to head with a Footscan pressure plate, the lab reference. Heel to toe, STRIV’s center of pressure follows the plate’s.
52 feet · 394 steps- Front-to-back center of pressure
- r0.998
- Side-to-side center of pressure
- r0.88
- Pressure distribution, heel to toeShare of pressure across 7 equal zones
- r0.89
52 feet, running in own shoes, thin shoes and socks. Both measures use the middle 80% of each stance and are scored on 394 steps, with the plate and insole axes aligned per foot.